EMD Millipore Corporation recalls Lugol s Iodine Solution 500MLDue to the incorrect packaging utilized with lot 3124, there' s an improper seal between the cap and bottle. With an improper seal the Assay (current specification of 24.8 to 27.2 mL) is expected to continue to drop over time as iodine vapors continue to dissipate from the bottles.
-
Class IIofficial
BioFire Diagnostics, LLC recalls BIOFIRE Blood Culture Identification 2 (BCID2) PanelRoutine quality control testing of affected blood culture identification panel with affected control panel may need to be amended, due to control panel/instrument characteristic changes, that lead C. tropicalis to be amplified in some portion of test runs, but then it's reported as "Not Detected" because the assay amplicon Tm is outside the acceptable limit assigned for calling positive results.
-
Class IIofficial
Bebig Isotopentechnik Gmbh recalls SagiPlan 2.2Due to software malfunction, numerical values may be rounded resulting in inaccurate measurements.
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls ARTIS icono (ceiling configuration)Potential hardware issue for ARTIS icono ceiling system: collision sensor may falsely signal non-existing collisions for up to half an hour when the system is switched off for a longer time (e.g., overnight) and switched on. During this time, system movement is only possible with the "override" function with very slow speed and without further collision protection. This may lead to temporary unavailability of the system for diagnostic and therapeutic usage.
-
Class IIofficial
Aizu Olympus Co., Ltd. recalls CF-HQ190L EVIS EXERA III COLONOVIDEOSCOPESpecific devices may have an improperly repaired light guide connector due to missing adhesive to secure the suction channel mouthpiece to the scope connector. Issue may result in loss of image, decreased suction, or impact the ability to effectively reprocess the device. This in turn may lead to unanticipated delay to treatment, or patient exposure to contaminated material and subsequent infection.
-
Class IIofficial
Aizu Olympus Co., Ltd. recalls GIF-H190Specific devices may have an improperly repaired light guide connector due to missing adhesive to secure the suction channel mouthpiece to the scope connector. Issue may result in loss of image, decreased suction, or impact the ability to effectively reprocess the device. This in turn may lead to unanticipated delay to treatment, or patient exposure to contaminated material and subsequent infection.
-
Class IIofficial
Aizu Olympus Co., Ltd. recalls GIF-HQ190Specific devices may have an improperly repaired light guide connector due to missing adhesive to secure the suction channel mouthpiece to the scope connector. Issue may result in loss of image, decreased suction, or impact the ability to effectively reprocess the device. This in turn may lead to unanticipated delay to treatment, or patient exposure to contaminated material and subsequent infection.
-
Class IIofficial
Aizu Olympus Co., Ltd. recalls PCF-H190DLSpecific devices may have an improperly repaired light guide connector due to missing adhesive to secure the suction channel mouthpiece to the scope connector. Issue may result in loss of image, decreased suction, or impact the ability to effectively reprocess the device. This in turn may lead to unanticipated delay to treatment, or patient exposure to contaminated material and subsequent infection.
-
Class IIofficial
Aizu Olympus Co., Ltd. recalls PCF-H190LSpecific devices may have an improperly repaired light guide connector due to missing adhesive to secure the suction channel mouthpiece to the scope connector. Issue may result in loss of image, decreased suction, or impact the ability to effectively reprocess the device. This in turn may lead to unanticipated delay to treatment, or patient exposure to contaminated material and subsequent infection.
-
Class Iofficial
MEDLINE INDUSTRIES, LP - Northfield recalls Tracheostomy care kits: (1) Medline Tracheostomy Clean and Care Tray with 14 Fr Suction CatheterThe component tracheostomy brush has a sharp edge at the tip that can potentially puncture tubing during use and/or cause user injury. Additionally, there have been incidents of the tracheostomy brush bristles detaching prior to use in the packaging and/or during use.
-
Class Iofficial
MEDLINE INDUSTRIES, LP - Northfield recalls (1) Nursing Skills kitThe component tracheostomy brush has a sharp edge at the tip that can potentially puncture tubing during use and/or cause user injury. Additionally, there have been incidents of the tracheostomy brush bristles detaching prior to use in the packaging and/or during use.
-
Class Iofficial
MEDLINE INDUSTRIES, LP - Northfield recalls (1) Nurse KitThe component tracheostomy brush has a sharp edge at the tip that can potentially puncture tubing during use and/or cause user injury. Additionally, there have been incidents of the tracheostomy brush bristles detaching prior to use in the packaging and/or during use.
-
Class IIofficial
Zimmer Biomet Spine Inc. recalls Vitality Polyaxial ScrewSpinal fixation system 7.5mmx 50mm screws are packaged with the correct labeling, but the screws themselves were laser marked with an incorrect size description of 7.5 x 40; use of incorrect screws could lead to screw perforation and indention of vascular or other structures.
-
Class Iofficial
Baxter Healthcare Corporation recalls Novum IQ Syringe infusion systemBaxter is issuing an Urgent Medical Device Correction for the Novum IQ Syringe Pump. Baxter identified that after multiple downstream occlusion alarms, the pump may display an Infusion Complete alarm even though uninfused fluid remains in the syringe.
-
Class IIofficial
Avanos Medical, Inc. recalls AVANOS BALLARD Closed Suction System for AdultsCertain lots of BALLARD Closed Suction System for Adults, 14 Fr, DSE, MDI were labeled as not being made with Di(2-ethylhexyl) phthalate (DEHP); however, Avanos has determined that the tubing components of the product did contain DEHP.
-
Class IIofficial
Breas Medical, Inc. recalls with Notch 22M/15F sold as part of Breas Medical MPV circuit 2 pc Kits: 006093Produced with unintended open slits on the side of the mouthpiece body, may result in insufficient ventilation in spontaneous breathing patients using MPV-VCV and MPV-PCV modes of ventilation
-
Class Iofficial
Olympus Corporation of the Americas recalls BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Model Numbers BF-1T150There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the endoscope.
-
Class Iofficial
Olympus Corporation of the Americas recalls EVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Model Number Q180-ACThere is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the endoscope.
-
Class Iofficial
Olympus Corporation of the Americas recalls EVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Model Numbers BF-1T180There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the endoscope.
-
Class Iofficial
Olympus Corporation of the Americas recalls EVIS EXERA III BRONCHOVIDEOSCOPE Model Numbers BF-1TH190There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the endoscope.
-
Class Iofficial
Olympus Corporation of the Americas recalls EVIS EXERA b BRONCHOVIDEOSCOPE OLYMPUS Model Number BF-XT190There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the endoscope.
-
Class Iofficial
Olympus Corporation of the Americas recalls EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS Model Number BF-P190There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the endoscope.
-
Class IIofficial
Change Healthcare Canada Company recalls Change Healthcare Radiology SolutionsChange Healthcare has identified a software issue where Change Healthcare Radiology Solutions 14.0, 14.1, and 14.2 may not update records in external applications.
-
Class IIofficial
Maquet Medical Systems USA recalls Cardiohelp-iIt was found that the fixation (weld) employed by the supplier of the nut bolt to the slide rail plate was insufficient. The deficient weld allows the nut bolt to break off from the slide rail plate with slight manual force.
-
Class IIofficial
Medtronic Inc recalls Medtronic Mo.Ma Ultra Proximal Cerebral Protection DeviceThe Mo.Ma Ultra device manifold is labelled with 2 stickers, an inferior and superior sticker, one on each side of the manifold. These stickers label the common carotid artery (CCA) proximal balloon and external carotid artery (ECA) distal balloon inflation and deflation ports. Both sides of the complaint units manifold were labelled with the superior label leading to an incorrect identification of the proximal and distal inflation /deflation ports when viewing the devices from the inferior side.
-
Class IIofficial
Medtronic Inc recalls Medtronic Mo.Ma Ultra Proximal Cerebral Protection DeviceThe Mo.Ma Ultra device manifold is labelled with 2 stickers, an inferior and superior sticker, one on each side of the manifold. These stickers label the common carotid artery (CCA) proximal balloon and external carotid artery (ECA) distal balloon inflation and deflation ports. Both sides of the complaint units manifold were labelled with the superior label leading to an incorrect identification of the proximal and distal inflation /deflation ports when viewing the devices from the inferior side.
-
Class IIofficial
Medtronic Inc recalls Medtronic Mo.Ma Ultra Proximal Cerebral Protection DeviceThe Mo.Ma Ultra device manifold is labelled with 2 stickers, an inferior and superior sticker, one on each side of the manifold. These stickers label the common carotid artery (CCA) proximal balloon and external carotid artery (ECA) distal balloon inflation and deflation ports. Both sides of the complaint units manifold were labelled with the superior label leading to an incorrect identification of the proximal and distal inflation /deflation ports when viewing the devices from the inferior side.
-
Class Iofficial
Unomedical A/S recalls VariSoft infusion setThe Varisoft infusion set is used in conjunction with tandem insulin pumps. The defect reported was damage to the connector piece during manufacturing. This damage results in the connector becoming detached more easily from the insulin set than expected, requiring less force to disconnect than intended. Disconnection, when not detected, for example during sleep, leads to missed basal dosing. which in turn can result in elevated blood glucose and ketone level (nocturnal hyperglycemia). Diabetic ketoacidosis.
-
Class Iofficial
Unomedical A/S recalls VariSoft infusion setThe Varisoft infusion set is used in conjunction with tandem insulin pumps. The defect reported was damage to the connector piece during manufacturing. This damage results in the connector becoming detached more easily from the insulin set than expected, requiring less force to disconnect than intended. Disconnection, when not detected, for example during sleep, leads to missed basal dosing. which in turn can result in elevated blood glucose and ketone level (nocturnal hyperglycemia). Diabetic ketoacidosis.hyperglycemia). Diabetic ketoacidosis.
-
Class Iofficial
Unomedical A/S recalls VariSoft infusion setThe Varisoft infusion set is used in conjunction with tandem insulin pumps. The defect reported was damage to the connector piece during manufacturing. This damage results in the connector becoming detached more easily from the insulin set than expected, requiring less force to disconnect than intended. Disconnection, when not detected, for example during sleep, leads to missed basal dosing. which in turn can result in elevated blood glucose and ketone level (nocturnal hyperglycemia). Diabetic ketoacidosis.
-
Class IIofficial
Becton Dickinson & Company recalls BD Alaris Pump infusion sets (Product NameAffected infusion sets labeled as free of DEHP actually do contain di(2-ethylhexyl) phthalate (DEHP).
-
Class IIofficial
Becton Dickinson & Company recalls BD gravity sets (Product NameAffected infusion sets labeled as free of DEHP actually do contain di(2-ethylhexyl) phthalate (DEHP).
-
Class IIofficial
Becton Dickinson & Company recalls BD Extension Sets (Product NameAffected infusion sets labeled as free of DEHP actually do contain di(2-ethylhexyl) phthalate (DEHP).
-
Class IIofficial
SunMed Holdings, LLC recalls VentlabA backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
-
Class IIofficial
SunMed Holdings, LLC recalls VentlabA backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
-
Class IIofficial
SunMed Holdings, LLC recalls VentlabA backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
-
Class IIofficial
SunMed Holdings, LLC recalls VentlabA backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
-
Class IIofficial
SunMed Holdings, LLC recalls VentlabA backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
-
Class IIofficial
SunMed Holdings, LLC recalls VentlabA backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
-
Class IIofficial
SunMed Holdings, LLC recalls VentlabA backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
-
Class IIofficial
SunMed Holdings, LLC recalls Curaplex manual resuscitators with integrated manometerA backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
-
Class IIofficial
SunMed Holdings, LLC recalls Medline manual resuscitators with integrated manometerA backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
-
Class IIofficial
Randox Laboratories Ltd. recalls RANDOX Total Bilirubin.Randox has had reports of elevated patient results using Plasma (lithium heparin) samples with Total Bilirubin reagent (TBil) Vanadate Oxidation Method.
-
Class IIofficial
Mc3 Inc recalls NautilusTM ECMO Oxygenator with Balance Biosurface-used in extracorporeal life support procedures to oxygenateMay have increased water path restriction within the heat exchanger, leading to degraded heat exchange performance during use. The reduction of water flow and reduced heat exchange performance may be experienced immediately upon priming or may worsen over the duration of use. The potential harm as stated in the IFU includes hypothermia due to inadequate heat transfer
-
Class IIofficial
Mc3 Inc recalls NautilusTM Smart ECMO Module with Balance Biosurface-used in extracorporeal life support procedures to oxygenaMay have increased water path restriction within the heat exchanger, leading to degraded heat exchange performance during use. The reduction of water flow and reduced heat exchange performance may be experienced immediately upon priming or may worsen over the duration of use. The potential harm as stated in the IFU includes hypothermia due to inadequate heat transfer
-
Class Iofficial
Maquet Medical Systems USA recalls CARDIOHELP Emergency DriveProduct removal due to possible blocking or impairment of the CARDIOHELP Emergency Drive. This issue can lead to the patient being exposed to inappropriate low blood flow. Potential associated harms are ischemia and hypoxia.
-
Class IIofficial
Abiomed, Inc. recalls Automated Impella Controller (AIC)Pump not detected as connected to controller due to software issue.
-
Class IIofficial
Abiomed, Inc. recalls Automated Impella Controller (AIC)Pump not detected as connected to controller due to software issue.
-
Class IIofficial
Fresenius Medical Care Holdings, Inc. recalls Stay Safe CapUpdating the labeling to include additional warnings and contraindications for potential side effects.
-
Class IIofficial
Ortho-Clinical Diagnostics, Inc. recalls VITROS 3600 Immunodiagnostic System Product Code 6802783 and Refurbished Product Code 6802914 Running SoftwareDuring planned monitoring of data post-release of Software Version 3.8.0 an increased occurrence of software issues relating to unresponsive subsystems, unresponsive user interface (sometimes referred to as a screen freeze), and/or printer disconnections was identified. The consequence of these issues would be a delay in reporting results due to the required system shutdowns to recover the system. This delay in reporting results could occur during any assay, including which the speed of the result is deemed critical or STAT and could result in a delay of treatment in a patient. Due to the release strategy of this software version, the user could not update the system but a field engineer had to manually install each affected system.